Forensicspedia

Categories
Live Forensic

Mass Spectrometry

⚛ Mass Spectrometry — How It Works

Ionisation · m/z Separation · Detection · Mass Spectrum

Sample
Introduction
Analyte
Ion Source
(EI / ESI)
IONISATION
M → M⁺• + e⁻
Mass Analyser
(Quadrupole)
+V
−V
Separates by m/z
Detector
(Electron Mult.)
EM Tube
Signal
Data
System
Mass Spectrum
Sample Molecules
Ionisation Energy (e⁻ beam)
Selected Ions (pass)
Rejected Ions (deflected)
Fragment Ions

1. Sample In

Analyte introduced as gas, liquid (ESI) or solid (MALDI) into the instrument under vacuum.

2. Ionisation

Molecules hit by electrons (EI) or charged spray (ESI) forming M⁺• radical cations.

3. Acceleration

Ions accelerated through electric potential into the mass analyser at high kinetic energy.

4. Separation

Quadrupole filters ions by m/z ratio — only selected masses pass through to detector.

5. Detection

Electron multiplier tube amplifies ion signal into measurable electrical current.

6. Spectrum

Software plots intensity vs m/z — each peak identifies a molecule or fragment ion.

▶ Live Mass Spectrum — Relative Intensity (%) vs m/z
% m/z → 100 75 50 25 15 29 43 57 71 85 99 114 M⁺ 43 85 M⁺ 114 100% M+1
📋 Step-by-Step Working of Mass Spectrometry
1

Sample Introduction

The analyte (the substance to be analysed) is introduced into the mass spectrometer through the inlet system. The entire instrument operates under high vacuum (10⁻⁵ to 10⁻⁸ torr) to prevent ions from colliding with air molecules. Samples can be introduced as a gas directly, as a liquid via a syringe or LC inlet (ESI), or as a solid using laser desorption (MALDI). The sample must be vaporised before it can be ionised.

2

Ionisation

Neutral molecules must be converted into ions before they can be separated by the mass analyser. There are several ionisation methods:

Electron Ionisation (EI): A beam of high-energy electrons (70 eV) bombards the sample vapour. An electron is knocked out of the molecule, creating a positively charged radical cation called the molecular ion (M⁺•). This is a “hard” ionisation method that causes significant fragmentation.

Electrospray Ionisation (ESI): The liquid sample is sprayed through a charged needle, creating multiply-charged ions. This is a “soft” method that keeps large molecules (like proteins) intact.

M + e⁻ → M⁺• + 2e⁻
3

Ion Acceleration

The newly formed positive ions are accelerated through a high electric potential (a few hundred to several thousand volts). This gives all ions of the same charge the same kinetic energy. Ions then enter the mass analyser as a focused beam. The key relationship here is:

KE = ½mv² = zV

Where m = mass, v = velocity, z = charge, V = accelerating voltage. Lighter ions move faster than heavier ones at the same kinetic energy, which allows them to be separated.

4

Mass Analysis — Separation by m/z

This is the heart of mass spectrometry. The mass analyser separates ions according to their mass-to-charge ratio (m/z). Different analysers use different principles:

Quadrupole: Four parallel rods with alternating DC and RF voltages. Only ions of a specific m/z have stable trajectories and pass through. Others are deflected and lost. By scanning the voltages, a full mass spectrum is recorded.

Time-of-Flight (TOF): Ions are accelerated over the same distance. Lighter ions arrive at the detector first. The time of flight is used to calculate m/z precisely.

Magnetic Sector: A magnetic field deflects ions — heavier ions are deflected less than lighter ones at the same velocity.

m/z = 2eV / v²
5

Ion Detection

Ions exiting the mass analyser strike the detector. The most common detector is the Electron Multiplier (EM). When a single ion hits the first dynode, it releases secondary electrons. These cascade through a series of dynodes, each releasing more electrons — amplifying the signal by a factor of 10⁵ to 10⁸. The final electrical current is proportional to the number of ions arriving, giving a relative intensity reading. Faraday cup detectors are also used for accurate quantitation of high-abundance ions.

Gain ≈ 10⁵ – 10⁸×
6

Data Processing — Mass Spectrum

The detector signals are sent to a data system which plots the mass spectrum — a graph of relative intensity (%) on the Y-axis versus m/z on the X-axis. Each vertical bar (peak) represents an ion of that particular m/z. The tallest peak is called the base peak (set to 100%). Key peaks to identify:

Molecular Ion Peak (M⁺): Appears at the highest m/z and corresponds to the intact molecule. Its m/z value gives the molecular weight of the compound directly.

Fragment Ion Peaks: Produced when the molecular ion breaks apart. The pattern of fragmentation is characteristic of the compound’s structure — like a fingerprint.

M+1 / M+2 Isotope Peaks: Small peaks one or two units above M⁺ due to naturally occurring ¹³C, ²H, ³⁷Cl, ⁸¹Br isotopes. Their relative heights help identify elemental composition.

Rel. Intensity = (Ion count / Base peak count) × 100%
⚡ Key Facts About Mass Spectrometry
  • Operates under high vacuum (prevents ion collisions with air)
  • Only ions (not neutral molecules) are detected
  • m/z = mass ÷ charge number of the ion
  • Base peak = most abundant ion (set to 100% intensity)
  • Molecular ion (M⁺) gives molecular weight directly
  • EI (70 eV) produces extensive fragmentation — good for structure ID
  • ESI is soft ionisation — preserves large biomolecules intact
  • Resolution = ability to distinguish two ions of similar m/z
  • Detection limit can be as low as femtomoles (10⁻¹⁵ mol)
  • Used in forensics, drug testing, proteomics, environmental analysis
  • LC-MS/MS = HPLC coupled to tandem mass spectrometry
  • HRMS (High Resolution MS) gives exact molecular formula
Categories
Live Forensic

HPLC (High-Performance Liquid Chromatography)

⚗ HPLC — How It Works

High-Performance Liquid Chromatography · Animated

Solvent
Reservoir
Mobile Phase
High-Pressure
Pump
PUMP
Sample
Injector
Separation Column
C18 Silica Stationary Phase
UV/Vis
Detector
λ = 254 nm
UV Cell
Waste /
Collector
Compound A — least retained (exits 1st)
Compound B — medium retained
Compound C — most retained (exits last)

1. Reservoir

Mobile phase stored & degassed before pumping.

2. Pump

Delivers solvent at constant high pressure up to 400 bar.

3. Injector

Sample loaded via loop valve into the solvent stream.

4. Column

Compounds separate based on affinity for stationary phase.

5. Detector

UV light detects each compound as it elutes from the column.

6. Output

Chromatogram plotted — each peak = one compound identified.

▶ Live Chromatogram Output
Absorbance Time → A B C